Cleaning the pressure transducer, Readjustment of the controller cvc 3000 – VACUUBRAND PC 3001 VARIO emission condenser Peltronic User Manual
Page 83
➨
Attention: Never use a pointed or sharp-edged tool
to clean the pressure transducer.
➨
Never touch the ceramic diaphragm of the pressure
transducer with hard objects.
➨
Fill the measurement chamber with a solvent (e.g.,
benzene) and allow sufficient cleaning time. Observe
all regulations concerning usage and disposal of sol-
vents!
➨
Drain the solvent and dispose of in accordance with
regulations. Repeat cleaning if necessary.
➨
Rinse the measurement chamber several times with al-
cohol in order to remove all solvent residues.
➨
Allow the pressure transducer to dry.
➨
Readjust the pressure transducer if necessary.
Readjustment of the controller CVC 3000
See section ”Readjustment of CVC 3000”, pg. 80.
Cleaning the pressure transducer
This manual is related to the following products: