PC 3001 TE VARIO:
• Comply with all applicable safety measures and require-
ments when using cryogenic coolants (e.g., dry ice).
Attention: Coolant media may cause severe frostbite
when coming in contact with skin!
• Prior to every use: Inspect the condenser for faults.
There must be no damage on the glass surface. Do not
use damaged components.
• Use only transport receptacles intended for coolants.
• Wear safety glasses and protective gloves.
• Do not clamp the covers of condensers or coolant re-
ceptacles. Allow pressure equalization between cool-
ant and atmosphere.
Do not start the pump if the pressure difference between
inlet and outlet exceeds 16 psi (1.1 bar) at maximum.
Prevent the backpressure of gases and the backflow of
condensates at the outlet.
Check the liquid level in the catchpots or in the collecting
flask regularly and drain condensate in time to prevent
overfilling. Install a level sensor (see ”Accessories”, pg
94) for monitoring, if necessary.
Provide appropriate protective measures to allow for the
possibility of failure and
malfunction. The protective mea-
sures must also allow for the requirements of the respec-
tive application.
The CVC 3000 controller is powered by a short-circuit-
proof wide-range power supply with integrated overload
protection.
A temperature sensor at the circuit board protects the mo-
tor: Current limitation in case the temperature at the circuit
board raises above 158°F (70°C). At temperatures above
185°F (85°C) the pump is switched off. In case of a motor
blockage (after 10 start-up attempts) the pump is switched
off.
Note: Only manual reset is possible. Disconnect the pump
NOTICE