Notice – VACUUBRAND PC 3001 VARIO emission condenser Peltronic User Manual
Page 50
page 50 of 120
• Ensure that the
coolant outlet hose is always free and
that it cannot get blocked.
• Maximum permissible coolant pressure at the exhaust
waste vapor condenser: 87 psi (6 bar) absolute
• Comply with the maximum permissible coolant pres-
sures of additional components in the coolant circuit
(e.g., coolant valve).
• We strongly recommend installing an optional coolant
valve (see ”Accessories”, pg. 94)
in the supply line of
the exhaust vapor condenser to save water and reduce
the risk of water spill.
• Avoid overpressure in the coolant circuit (e.g., caused
by blocked or kinked coolant hoses).
• Avoid any clogging of the exhaust vapor condenser
caused by deposits or frozen solvents. Clean if neces-
sary or use higher coolant temperatures.
In case of
condensation: Check the liquid level in both
catchpots (10) during operation. Check the liquid level reg-
ularly. Do not allow the catchpots to overfill. Drain catch-
pots in time to avoid overflow. Install a level sensor (see
”Accessories”, pg. 94) for monitoring, if necessary.
The maximum liquid level is at approximately 80% of the
total filling level to avoid problems when removing the
catchpots.
Permissible range of coolant temperature at the exhaust
waste vapor condenser:
5 °F to 68 °F (-15°C to +20°C)
Check hose connections prior to starting operation of the
cooling system.
Check coolant hoses regularly during operation.
Removing the catchpots:
Stop process.
Catchpot at outlet:
Remove joint clip. Remove catchpot and drain conden-
sate.
NOTICE