During operation, Notes concerning the operation of the, Peltronic emission condenser – VACUUBRAND PC 3001 VARIO emission condenser Peltronic User Manual
Page 46: Notice
page 46 of 120
Notes concerning the operation of the Peltronic emis-
sion condenser
Read the manual of the Peltronic emission condenser and
observe the instructions contained in that manual!
Assembling the Peltronic emission condenser:
➨
Assemble the Peltronic emission condenser
with two attachment screws (Phillips screw-
driver size 2) to the pumping unit (a).
➨
Assemble the hose connection between
pump outlet and condenser inlet (b).
➨
Slip connecting tube onto hose connection
of the Peltronic emission condenser. Secure
with hose clip.
➨
Connect the VACUU•BUS cable from the
CVC 3000 controller to the Peltronic emis-
sion condenser (c).
Notes concerning the operation of pumping units with
silencer
Attention: Dust-laden gases, deposits and condensed
solvent vapor can restrict air flow out the silencer. The re-
sultant back pressure can lead to damage of pump bear-
ings, diaphragms, and valves. Under those conditions, a
silencer must not be used.
During operation
➨
Vent and dispose of
potentially dangerous gases or
vapors at the outlet of the pump appropriately.
NOTICE
(a)
(b)
(c)
NOTICE