VACUUBRAND PC 3001 VARIO emission condenser Peltronic User Manual
Page 20
page 20 of 120
play flashes. Release pressure immediately! Risk of
bursting!
+
Comply with applicable regulations when disposing of
chemicals. Take into consideration that chemicals may
be contaminated. Take adequate precautions to pro-
tect people from the effects of dangerous substances
(chemicals, thermal decomposition products of fluo-
roelastomers). Use appropriate protective clothing and
safety goggles.
+
Use only
original manufacturer’s spare parts and
accessories. Otherwise the safety and performance of
the equipment, as well as the electromagnetic compat-
ibility of the equipment might be reduced.
The CE mark or the cTÜVus mark may be voided if not
using original manufacturer’s spare parts.
+
Interruption of the pump (e.g., due to power failure),
failure of connected components or of parts of the sup-
ply, or change in parameters must not be allowed to
lead to dangerous conditions. In case of a diaphragm
failure or in case of a leak in the manifold, pumped sub-
stances might be released into the environment or into
the pump housing or motor.
Comply with all notes regarding proper use of the pumps,
as well as operation and maintenance guidance.
+
The residual
leak rate of the equipment might render
possible an exchange of gas, albeit extremely slight,
between the environment and the vacuum system.
Adopt suitable measures to prevent contamination of
the pumped substances or the environment.
PC 3001 TE VARIO:
+
Cold surfaces can absorb a great amount of gas, which
may expand abruptly in case of warming. This may lead
to unmanageable overpressure in the system, creating
a risk of bursting!
+
Check the coolant level of the condenser at appropriate