• Ensure that the
coolant outlet pipe is always free and
that it cannot get blocked. If installing an optional cool-
ant valve, it must always be in the supply line of the
exhaust waste vapor condenser.
Keep a minimum distance of 8 in (20 cm) between the
cooling fan and surrounding items (e.g., housing, walls,
etc.). Clean ventilation slots if necessary to avoid a reduc-
tion of ventilation.
Do not place the pumping unit on soft surfaces (e.g., rub-
ber foam) during operation. This may reduce or block the
electronics’ air supply. Do not cover the pumping unit.
Use only hoses at the inlet and outlet of the pump with
an inner diameter at least as large as the diameter of the
pump’s tubing (to avoid overpressure at the outlet, and
reduction of pumping speed at the inlet).
Allow the equipment to equilibrate to ambient temperature
if you bring it from cold environment into a room prior to
operation. Notice if there is water condensation on cold
surfaces.
Secure coolant hoses at the hose nozzles (e.g., with hose
clamp) to prevent their accidental slipping.
Comply with all
applicable and relevant safety require-
ments (regulations and guidelines). Implement the re-
quired actions and adopt suitable safety measures.
Ambient conditions
➨
Do not reach for this product if it has fallen into liquid.
There is a risk of deadly electrical shock. Unplug the
system immediately.
+
Do not use this product in an area where it can fall or
be pulled into water or other liquids.
NOTICE