Notes regarding the (dry) ice condenser – VACUUBRAND PC 3001 VARIO emission condenser Peltronic User Manual
Page 43
•
Note: Install the hoses of the cooling system in a way
to avoid the flow / dripping of condensed water onto the
pumping unit (especially cables and electronic parts,
see also IP degree of protection, ”Technical data”, pg.
• Ensure that the
coolant outlet tubing is always unob-
structed and that it cannot get blocked.
• Maximum permissible coolant pressure at the exhaust
waste vapor condenser: 87 psi (6 bar) absolute. Outlet
flow must always be unhindered.
• Comply with the maximum permissible coolant pres-
sures of additional components in the coolant circuit
(e.g., coolant valve).
• Avoid overpressure in the coolant circuit (e.g., caused
by blocked or squeezed coolant hoses).
• Only install the optional coolant valve in the supply line
of the exhaust waste vapor condenser.
Notes regarding the (dry) ice condenser
(PC 3001 TE VARIO)
The
exhaust waste vapor condenser enables
an efficient condensation of the pumped vapors
at the outlet.
+
Close to 100% solvent recovery.
+
The isolation cover protects against glass
splinters in case of breaking, acts as thermal
insulation to avoid the condensation of hu-
midity and is intended to absorb shocks.
➨
Adopt suitable measures to prevent the formation of
explosive or flammable mixtures, use inert gas for vent-
ing if necessary.
➨
Prevent the discharge of dangerous gases and vapors
to the surrounding atmosphere. If appropriate, connect
the exhaust line to a suitable treatment system.
➨
Adopt suitable measures to prevent the release of dan-
gerous, toxic, explosive, corrosive, noxious or polluting
fluids, vapors and gases when disposing of conden-
sates.
outlet (gas!)
hose nozzle