Notice – VACUUBRAND PC 3001 VARIO emission condenser Peltronic User Manual
Page 42
page 42 of 120
➨
Assemble the catchpots at the inlet and at
the outlet using joint clips.
Exhaust waste vapor condenser
(PC3001 VARIO):
➨
Assemble the hose nozzles for coolant inlet
(11) and coolant outlet (12) tubing and for
the gas outlet (6) at the exhaust waste vapor
condenser.
The
exhaust waste vapor condenser enables an effi-
cient condensation of the pumped vapors at the outlet.
+
No backflow of condensates.
+
Controlled recovery of condensates.
+
Close to 100% solvent recovery.
+
The isolation cover protects against glass splinters in
case of breakage, acts as thermal insulation to avoid
condensation of humidity and is intended to absorb
shocks.
➨
Attach the tubing of the coolant circuit to the respective
hose nozzles (hose nozzles for tubing I.D. 1/4”-5/16”
(6-8 mm), see image) at the waste vapor condenser.
Check all hose connections prior to starting operation of
the cooling system.
Secure coolant hoses at the hose nozzles (e.g., with hose
clamps) to prevent their accidentally slipping off.
➨
Prevent the discharge of dangerous gases and vapors
to the surrounding atmosphere. If appropriate, connect
the exhaust line to a suitable treatment system.
+
Never block the gas outlet ((8) hose nozzle for tubing
I.D. 3/8” (10 mm)). The exhaust hose has always to be
unobstructed and without back pressure to enable an
unhindered discharge of gases and protect the pump
valves from damage.
11
12
6
NOTICE