Connecting components at the controller:
At the rear side of the controller are connections
for e.g., an optional coolant valve and/or an ex-
ternal pressure transducer and/or an external
venting valve, as well as a venting connection
and the serial interface RS 232C.
- Pull the controller out of the housing. Be careful not to
disconnect the measuring connection (PTFE tube)!
- Insert the cable of the component to be connected un-
derneath the housing and plug into the controller. Do
not apply off-axis forces when assembling or removing
plug connections! Observe correct orientation of the
plug.
- Insert controller into the housing.
Separator at inlet and exhaust waste vapor condens-
er:
Assembling the hose nozzle with union nut:
➨
Take the hose nozzle with attached compression fer-
rule and union nut out of the catchpot and put onto inlet
connection.
➨
Tighten the union nut by hand until you can feel the
stop. Then tighten an additional 1/4 rotation with an
open-ended wrench (size 17mm) for final installation.
Catchpots:
The catchpot at the inlet protects against drop-
lets and particles from entering the pump.
+
Enhances lifetimes of diaphragms and
valves.
+
Improves vacuum performance in applica-
tions with condensable vapors.
Both catchpots are coated with a protective lay-
er to protect against shattering in case of break-
age or implosion.
catchpot at
the inlet (10)
catchpot at the outlet (10)