beautypg.com

9 test probe calibration 39, 1 calibrating a single probe 39, 2 dry-well characteristics 39 – ThermoWorks 9141 HART SCIENTIFIC FIELD DRY-WELL TEMPERATURE CALIBRATORS User Manual

Page 41: 1 heating and cooling rates 39, 2 stabilization and accuracy 39, Test probe calibration, Calibrating a single probe, Dry-well characteristics, Heating and cooling rates, Stabilization and accuracy

background image

9

Test Probe Calibration

For optimum accuracy and stability, allow the calibrator to warm up for 10
minutes after power-up and then allow adequate stabilization time after reach-
ing the set-point temperature. After completing operation of the calibrator, al-
low the well to cool by setting the temperature to 100°C for one-half hour
before switching the power off.

9.1

Calibrating a Single Probe

Insert the probe to be calibrated into the well of the dry-well calibrator. The
probe should fit snugly into the calibrator probe sleeve yet should not be so
tight that it cannot be easily removed. Avoid any dirt or grit that may cause the
probe to jam into the sleeve. Best results are obtained with the probe inserted to
the full depth of the well. Once the probe is inserted into the well, allow ade-
quate stabilization time to allow the test probe temperature to settle as de-
scribed above. Once the probe has settled to the temperature of the well, it may
be compared to the calibrator display temperature. The display temperature
should be stable to within 0.1°C degree for best results.

CAUTION: Never introduce any foreign material into the probe hole of

the insert. Fluids etc. can leak into the calibrator causing damage to the
calibrator or binding and damage to your probe.

9.2

Dry-well Characteristics

There is a temperature gradient vertically in the test well. The heater has been
applied to the block in such a way as to compensate for nominal heat losses out
of the top of the dry-well. However, actual heat losses will vary with design of
the thermometer probes inserted into the calibrator and the temperature. For
best results, insert probe to full depth of well.

9.2.1

Heating and Cooling Rates

Figures 7 and 8 show typical heating cooling rates of the 9141 dry-well
calibrator.

DO NOT remove inserts when heating or when the unit is hot.

9.2.2

Stabilization and Accuracy

The stabilization time of the dry-well calibrator will depend on the conditions
and temperatures involved. Typically the test well will be stable to 0.1°C within
5 minutes of reaching the set-point temperature as indicated by the display. Ul-
timate stability will be achieved 10 to 20 minutes after reaching the set
temperature.

39

9 Test Probe Calibration

Calibrating a Single Probe