Table 2 9141 communication commands continued 37, Table 2, 9141 communication commands continued – ThermoWorks 9141 HART SCIENTIFIC FIELD DRY-WELL TEMPERATURE CALIBRATORS User Manual

Page 40

background image

37

8 Digital Communication Interface

Interface Commands

Command Description

Command
Format

Command
Example

Returned

Returned
Example

Acceptable
Values

Serial Interface Menu

Read serial sample setting

sa[mple]

sa

sa: 9

sa: 1

Set serial sampling setting to

n

seconds

sa[mple]=n

sa=0

0 to 999

Set serial duplex mode:

du[plex]=f[ull]/h[alf]

FULL or HALF

Set serial duplex mode to full

du[plex]=f[ull]

du=f

Set serial duplex mode to half

du[plex]=h[alf]

du=h

Set serial linefeed mode:

lf[eed]=on/of[f]

ON or OFF

Set serial linefeed mode to on

lf[eed]=on

lf=on

Set serial linefeed mode to off

lf[eed]=of[f]

lf=of

Miscellaneous

Read firmware version number

*ver[sion]

*ver

ver.9999,9.99

ver.9141,1.21

Read structure of all commands

h[elp]

h

list of commands

Read ALL operating parameters

all

all

list of parameters

Legend:

[] Optional Command data

{} Returns either information

n Numeric data supplied by user

9 Numeric data returned to user

x Character data returned to user

Note:

When DUPLEX is set to FULL and a command is sent to READ, the command is returned followed by a
carriage return and linefeed. Then the value is returned as indicated in the RETURNED column.

9141 communication commands continued