ture. The factory default and maximum are set to 650°C. For safety, a user can
adjust the HL down so the maximum temperature set-point is restricted.

HL

High Limit parameter

Press “SET” to enable adjustment of HL

H=650

Current HL setting

Adjust the HL parameter using “UP” or “DOWN”.

H=400

New HL setting

Pres “SET” to accept the new temperature limit.

7.11

Serial Interface Parameters

The serial RS-232 interface parameters menu is indicated by,

SErIAL

Serial RS-232 interface parameters menu

The serial interface parameters menu contains parameters which determine the
operation of the serial interface. These controls only apply to instruments fitted
with the serial interface. The parameters in the menu are — BAUD rate, sample
period, duplex mode, and linefeed.

7.11.1

Baud Rate

The baud rate is the first parameter in the menu. The baud rate setting deter-
mines the serial communications transmission rate.

The baud rate parameter is indicated by,

bAUd

Serial BAUD rate parameter

Press “SET” to choose to set the baud rate. The current baud rate value will
then be displayed.

2400 b

Current baud rate

The baud rate of the serial communications may be programmed to 300, 600,
1200, 2400, 4800, or 9600 baud. Use “UP” or “DOWN” to change the baud
rate value.

4800 b

New baud rate

9141 Dry-well Calibrator
User’s Guide

30