does not prevent all heat from reaching the hub. Be careful when remov-
ing the heat shield or the probe from the well.

DO NOT apply any type of voltage to the display hold terminals. Apply-

ing a voltage to the terminals may cause damage to the controller.

DO NOT use fluids to clean out the well. Fluids could leak into electron-

ics and damage the instrument.

Never introduce any foreign material into the probe hole of the insert.

Fluids, etc. can leak into the instrument causing damage.

DO NOT change the values of the calibration constants from the factory

set values. The correct setting of these parameters is important to the
safety and proper operation of the calibrator.

DO NOT slam the probe sheath in to the well. This type of action can

cause a shock to the sensor and affect the calibration.

• The instrument and any thermometer probes used with it are sensitive in-

struments that can be easily damaged. Always handle these devices with
care. DO NOT allow them to be dropped, struck, stressed, or overheated.

• The Factory Reset Sequence (see Section 12, Troubleshooting) should be

performed only by authorized personnel if no other action is successful in
correcting a malfunction. You must have a copy of the most recent Report
of Calibration to restore the calibration parameters.