2 set-point value 23, 3 temperature scale units 23, Set-point value – ThermoWorks 9141 HART SCIENTIFIC FIELD DRY-WELL TEMPERATURE CALIBRATORS User Manual
Page 26: Temperature scale units

100.0 C
Well temperature in degrees Celsius
S
Access set-point memory
1. 100.
Set-point memory 1, 100°C currently used
To change to another set-point memory press “UP” or “DOWN”.
4. 300.
New set-point memory 4, 300°C
Press “SET” to accept the new selection and access the set-point value.
S
Accept selected set-point memory
7.2.2
Set-point Value
The set-point value may be adjusted after selecting the set-point memory and
pressing “SET”.
4 200.
Set-point 4 value in°C
If the set-point value is correct then press “EXIT” to resume displaying the well
temperature. To change the set-point values, press “SET” and then press “UP”
or “DOWN” to adjust the set-point value.
220.
New set-point value
When the desired set-point value is reached press “SET” to accept the new
value and access the temperature scale units selection. If “EXIT” is pressed in-
stead then any changes made to the set-point will be ignored.
S
Accept new set-point value
7.2.3
Temperature Scale Units
The temperature scale units of the controller maybe set by the user to degrees
Celsius (°C) or Fahrenheit (°F). The units are used in displaying the well tem-
perature, set-point, and proportional band.
Press “SET” after adjusting the set-point value to change display units.
Un= C
Scale units currently selected
Press “UP” or “DOWN” to change the units.
23
7 Controller Operation
Temperature Set-point