9141 Dry-well Calibrator
User’s Guide

36

Command Description

Command
Format

Command
Example

Returned

Returned
Example

Acceptable
Values

Display Temperature

Read current set-point

s[etpoint]

s

set: 9999.99 {C or F}

set: 150.00 C

Set current set-point to n

s[etpoint]=n

s=350

Instrument
Range

Read temperature

t[emperature]

t

t: 9999.9 {C or F}

t: 55.6 C

Read temperature units

u[nits]

u

u: x

u: C

Set temperature units:

u[nits]=c/f

C or F

Set temperature units to Celsius

u[nits]=c

u=c

Set temperature units to
Fahrenheit

u[nits]=f

u=f

Read scan mode

sc[an]

sc

sc: {ON or OFF}

sc: ON

Set scan mode

sc[an]=on/off

sc=on

ON or OFF

Read scan rate

sr[ate]

sr

srat: 99.9 {C or F}/min

srat:12.4 C/min

Set scan rate

sr[ate]=n

sr=1.1

.1 to 99.9

Read hold

ho[ld]

ho

ho: open/closed, 99.9 {C
or F}

ho: open, 30.5C

Secondary Menu

Read proportional band setting

pr[opband]

pr

pb: 999.9

pb: 15.9

Set proportional band to

n

pr[opband]=n

pr=8.83

Depends on
Configuration

Read heater power

(duty cycle)

po[wer]

po

po: 999.9

po: 1.3

Configuration Menu

Calibration Parameters Menu

Read R0 calibration parameter

r[0]

r

r0: 999.999

r0: 100.578

Set R0 calibration parameter to

n r[0]=n

r=100.324

98.0 to 104.9

Read ALPHA calibration
parameter

al[pha]

al

al: 9.9999999

al: 0.0038573

Set ALPHA calibration parameter
to

n

al[pha]=n

al=0.0038433

.002 to .006

Read DELTA calibration
parameter

de[lta]

de

de: 9.9999

de: 1.507

Set DELTA calibration parameter de[lta]=n

de=1.3742

0–3.0

Operating Parameters Menu

Read High Limit

hl

hl

hl: 999

hl: 600

Set High Limit

hl=

n

hl=600

100–650

Table 2 9141 controller communications commands