Page 31: Calibration parameters

background image

Press “SET” and “EXIT” to enter the secondary menu and show the heater
power. Then press “SET” to access the proportional band.

S

+

E

Access heater power in secondary menu

SEC

Flashes

for secondary menu and then displays the

heater power

12.0 P

Heater power in percent

S

Access proportional band

4.1

Proportional band setting

ProP

Flashes

and then displays the value

To change the proportional band press “UP” or “DOWN”.

10.0

New proportional band setting

To accept the new setting press “SET”. Press “EXIT” to continue without stor-
ing the new value.

S

Accept the new proportional band setting

7.8

Controller Configuration

The controller has a number of configuration and operating options and calibra-
tion parameters which are programmable via the front panel. These are ac-
cessed from the secondary menu after the proportional band function by
pressing “SET”. Pressing “SET” again enters the first of three sets of configu-
ration parameters — calibration parameters, operating parameters and serial in-
terface parameters. The sets are selected using the “UP” and “DOWN” keys
and then pressing “SET”. (See

Figure 5

on page 22.)

7.9

Calibration Parameters

The operator of the instrument controller has access to a number of the calibra-
tion constants namely R0, ALPHA, and DELTA. These values are set at the
factory and must not be altered. The correct values are important to the accu-
racy and proper and safe operation of the instrument. Access to these parame-
ters is available to the user so that in the event that the controller memory fails

9141 Dry-well Calibrator
User’s Guide

28