Sterling Scale 825 User Manual
Page 14

_____________________________________________________________________________________________________________
Revision 4.4 Copyright ©2004Sterling Scale Co., Inc. March 2005
Web Site:
www.sterlingscale.com
E-mail: [email protected]
9.
The display will repeatedly flash “CLEAR” and “SCALE”. When the scale is cleared of all
weight, press the PRINT key.
10.
The display shows “STEP A8”. The indicator will count down to “STEP A0” and then display
one of the following messages: “ENT LB”, “ENT KG”, or “ENT G”. Lightly press the PRINT
key to see the previous calibration weight used.
11.
Use the number keys to select the desired calibration weight. Once the desired calibration
weight is entered, place the weight on the scale and press the PRINT key.
12.
The display shows “STEP B8”. The indicator will count down to “STEP B0” and ready to exit
Calibration Setup.
13.
The indicator is now at the Fine Span Adjust parameter, SPN 0.0.
14. Hit the RESET key twice to exit parameter set up.
PARAMETER SETUP AND CALIBRATION
Page 12