L
IMIT
Perform the following steps to set up the Over Temperature Limit system for use.
1. If you have not done so already, turn the Set Over Temperature Limit control dial clockwise to
the maximum position. This allows the Set Temperature control to stabilize.
2. Then turn the Over Temperature Limit control dial counterclockwise until the Over Temp Limit
Activated light illuminates.
3. Slowly turn the Over Temperature Limit control dial clockwise until the light turns off.
4. This adjusts the Over Temperature Limit control to approximately 1
˚C above the temperature
configured by the Set Temperature control.
S
ILENCING THE
D
OOR
A
JAR
A
LARM
If the door alarm needs to be silenced, close the door, which will reset the timer. For prolonged periods
with the door open, the unit should be switched to the OFF position. If the door alarm continues to sound
when the door is fully closed, the position of the black tab-like door alarm sensor, mounted on the front
left side of the door frame overhang, can be adjusted using a screwdriver. Loosen the screws just enough
to slide the sensor and adjust. Do not remove the screws.
L
OADING THE
I
NCUBATOR
Place items on the shelves inside the incubator chamber as evenly spaced as possible. Good spacing
allows for maximum air circulation and a higher degree of temperature uniformity.
A
TTACHING EQUIPMENT TO THE
I
NTERIOR
A
CCESSORY
O
UTLET
The incubator has a 1A (maximum) accessory outlet located inside the chamber. The power switch on the
front panel controls power to the accessory outlet. This outlet can power equipment such as magnetic
stirrers, rockers, etc. Do not attach any equipment drawing more than 1A to this outlet.
Accessory equipment may produce additional heat. This heat could affect the temperature range and
uniformity of this incubator. If you are using the interior accessory outlet, you should check that the
incubator operates within required temperatures when the accessory equipment is installed and
operating.