OPERATION (CONTINUED)

End of procedure

Calibrating the CO

2

Display (Continued)


Reference

6. Compare the reference CO

2

analyzer reading to the reading of the

incubator CO

2

display.

a. If the reference device and the incubator display

readings are the same or the difference falls within the
range of your laboratory protocol, the incubator is now
calibrated for CO

2

.





7.

If the readings falls outside your laboratory protocol’s range, enter a
new adjustment offset,

repeat steps 3

– 6

, starting on page 29.

a. Repeat for a total of three attempts.



8. If the gas CO

2

concentration readings of the incubator and the

reference device still fall outside your laboratory protocol after three
calibration attempts, contact

Sheldon Technical Support

for

assistance.



Note
: Three attempts
may be required to
sucessfully calibrate a
unit that is more than 2%
out of calibration.