Lifting, Leveling, Access port stopper – Shellab SCO31 User Manual
Page 12: Installation (continued)

12 |
P a g e
INSTALLATION (CONTINUED)
L
IFTING
A SCO Incubator should only be lifted by its bottom surfaces using proper heavy lifting machinery
such as, a forklift or pallet jack. Handles and knobs are inadequate for lifting or stabilization. The
unit should be completely restrained from tipping during lifting. Transporting the unit while lifted is
not recommended and may be hazardous. Remove all moving parts, such as shelves and trays,
and secure the door in the closed position prior to lifting the unit.
Do not attempt to move the unit while in operation or before the unit has cooled.
L
EVELING
The SCO Incubator must be level and stable for safe operation. Each incubator ships with four
leveling feet. Insert one leveling foot into each of the four holes in the bottom corners of the
workstation. Adjust the foot at each corner until the workstation stands level and solid without
rocking. To raise a foot, turn it in a counterclockwise direction; to lower a foot, turn it in a clockwise
direction.
A
CCESS
P
ORT
S
TOPPER
The SCO58 Incubator is provided with an access port located on the right side of the
unit. The incubator is shipped with one (1) rubber access port stopper. The stopper
should come installed in the port inside the incubation chamber. The stopper should
always be installed in the chamber to obtain the best temperature uniformity and
prevent condensation from forming inside the port. A second stopper may be
installed on the outside of the unit to prevent dust from building up in the port, but is
not required.