Operation (continued) – Shellab SCO31 User Manual

Page 29

background image

29 |

P a g e

OPERATION (CONTINUED)

Procedure continued on next page

Calibrating the CO

2

Display (Continued)



3. If there is a difference between the readings, and that
difference

falls outside your laboratory protocol’s acceptable

range, place the incubator into calibration mode by pressing and
holding the CO

2

Control Up and Down arrow buttons.

a. The incubator CO

2

display will flash the letters “CP”, and

then show a blinking adjustable CO

2

concentration

value. Release the arrow buttons.

b. Use the Up and Down arrows to adjust the blinking

display value to match the CO

2

analyzer reading.

c. When you have matched the display value to the reading of the

CO

2

analyzer, wait for five (5) seconds.




4. The incubator display will cease flashing. The new calibrated display

value is now stored in the unit. The display will revert to showing the
current CO

2

level in the chamber, and the incubator will adjust the

chamber CO

2

level to match the current set point.


5. Allow the incubator sit for one (1) hour undisturbed to
stabilize after it has achieved the CO

2

set point with the

corrected display value. Failure to wait until the unit is fully
stabilized will result in an inaccurate reading during the next
step.