Operation (continued) – Shellab SCO31 User Manual
Page 29

29 |
P a g e
OPERATION (CONTINUED)
Procedure continued on next page
Calibrating the CO
2
Display (Continued)
3. If there is a difference between the readings, and that
difference
falls outside your laboratory protocol’s acceptable
range, place the incubator into calibration mode by pressing and
holding the CO
2
Control Up and Down arrow buttons.
a. The incubator CO
2
display will flash the letters “CP”, and
then show a blinking adjustable CO
2
concentration
value. Release the arrow buttons.
b. Use the Up and Down arrows to adjust the blinking
display value to match the CO
2
analyzer reading.
c. When you have matched the display value to the reading of the
CO
2
analyzer, wait for five (5) seconds.
4. The incubator display will cease flashing. The new calibrated display
value is now stored in the unit. The display will revert to showing the
current CO
2
level in the chamber, and the incubator will adjust the
chamber CO
2
level to match the current set point.
5. Allow the incubator sit for one (1) hour undisturbed to
stabilize after it has achieved the CO
2
set point with the
corrected display value. Failure to wait until the unit is fully
stabilized will result in an inaccurate reading during the next
step.