OPERATION (CONTINUED)

Temperature Calibration Continued

10. If the temperature readings of the incubator and the reference device still fall outside your

laboratory protocol after three calibration attempts, contact

Sheldon Technical Support

for assistance.

a. Three calibration attempts may be required to successfully calibrate units that

are ±2°C or more out of calibration.

End of Procedure

S

ET THE

O

VER

T

EMPERATURE

L

IMIT

The incubator temperature must be stable running at your temperature set point for at least one
hour prior to setting the OTL. Perform the following steps to set up the Over Temperature Limit
system for normal use:

End of Procedure

1. If you have not done so already, turn the Set Over Temperature Limit

control dial clockwise to the maximum position. This allows the Set
Temperature control to stabilize.

2. Turn the Over Temperature Limit control dial counterclockwise until the

red Over Temp Limit Activated light illuminates.

3. Slowly turn the Over Temperature Limit control dial clockwise until the

Over Temp Limit Activated light turns off. Stop turning the control.

4. Leave the OTL control slightly above the activation point. This sets the

Over Temperature Limit set to approximately 1

˚C above the current

temperature set point.