19

SPeCIFICATIOnS

Guidance and manufacturer’s declaration - electromagnetic immunity

The table is intended for use in the electromagnetic environment specified below. The customer or the

user of the table should assure that it is used in such an environment.

Immunity Test

IeC 60601 test level Compliance level

electromagnetic

environment - guidance

Conducted rF

IeC 61000-4-6

radiated rF

IeC 61000-4-3

3 Vrms

150 kHz to 80 MHz

3 V/m

80 MHz to 2,5 GHz

3Vrms

3V/m

Portable and mobile rF communications equipment

should be used no closer to any part of the table, in-

cluding cables, than the recommended separation

distance calculated from the equation applicable

to the frequency of the transmitter.

Recommended separation distance

d = 1.2 √ P

d = 1.2 √ P

80 MHz to 800 MHz

d = 2.3 √ P

800 MHz to 2,5 GHz

Where P is the maximum output power rating of the

transmitter in watts (W) according to the transmitter

manufacturer and d is the recommended separa-

tion distance in meters (m).
Field strengths from fixed RF transmitters, as deter-

mined by an electromagnetic site survey,

a

should be

less than the compliance level in each frequency

range.

b

Interference may occur in the vicinity of equipment

marked with the following symbol:

NOTe 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTe 2 These guidelines may not apply in all situations. electromagnetic propagation is affected by absorption and

reflection from structures, objects and people.

a

Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile

radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accura-

cy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be

considered. If the measured field strength in the location in which the table is used exceeds the applicable RF com-

pliance level above, the table should be observed to verify normal operation. If abnormal performance is observed,

additional measures may be necessary, such as re-orienting or relocating the table

b

Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.