11

DIReCTIOnS fOR USe

PROCeDURe ChAIR fACe ReST CReSCenT

The Procedure Chair Face Rest Crescent is designed to be used for prone positioning. An Adjustable Head

Rest is required for this accessory.

T-RAIl ADAPTeR

The T-Rail Adapter can be attached at designated locations on one or both sides of the back rest sec-

tion, allowing the use of any t-rail accessory. Attach or remove using knobs. The T-Rail Adapter has a

Maximum Load Capacity of 150 lbs. (68 kg.)

POWeR OUTleT USe

The maximum amperage for the Power Outlet is 3 amps. Connecting equipment that exceeds this

rating will damage the system. Only connect low current devices. Do not connect an extension

cord or power strip to the Power Outlet. Do not touch equipment connected to the Power Outlet and the patient at the same

time. Any electrical accessories used with this table must comply with approved medical standards. Equipment relying only on

Basic Insulation (only one means of electrical protection for the patient / pratitioner) must not be used.

WARnInG

The Power Outlet can be used to connect low

power devices such as localized illumination,

computers, micro-controllers, etc. Exceeding the

maximum current will interrupt the power.