SPECIFICATIONS

Guidance and manufacturer’s declaration - electromagnetic immunity

The chair is intended for use in the electromagnetic environment specified below. The customer or the user of the

chair should assure that it is used in such an environment.

Immunity Test IEC 60601 test level

Compliance level

Electromagnetic environment - guidance

Conducted RF

IEC 61000-4-6

Radiated RF

IEC 61000-4-3

3 Vrms

150 kHz to 80 MHz

3 V/m

80 MHz to 2,5 GHz

3Vrms

3V/m

Portable and mobile RF communications equipment

should be used no closer to any part of the chair, includ-

ing cables, than the recommended separation distance

calculated from the equation applicable to the frequency

of the transmitter.

Recommended separation distance

d = 1.2 √ P

d = 1.2 √ P

80 MHz to 800 MHz

d = 2.3 √ P

800 MHz to 2,5 GHz

Where P is the maximum output power rating of the trans-

mitter in watts (W) according to the transmitter manufac-

turer and d is the recommended separation distance in

meters (m).
Field strengths from fixed RF transmitters, as deter-mined

by an electromagnetic site survey,

a

should be less than

the compliance level in each frequency range.

b

Interference may occur in the vicinity of equipment

marked with the following symbol:

NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection

from structures, objects and people.

a

Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios,

amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the

electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured

field strength in the location in which the chair is used exceeds the applicable RF compliance level above, the chair should be

observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-

orienting or relocating the table

b

Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 1 V/m.