SPECIFICATIONS

Guidance and manufacturer’s declaration - electromagnetic immunity

The chair is intended for use in the electromagnetic environment specified below. The customer or the user of the

chair should assure that it is used in such an environment.

Immunity Test

IEC 60601 test level

Compliance level

Electromagnetic

environment - guidance

Electrostatic discharge (ESD)

IEC 61000-4-2

±6 kV contact

±8 kV air

±6 kV contact

±8 kV air

Floors should be wood, con-

crete or ceramic tile. If floors

are covered with synthetic

material, the relative humidity

should be at least 30 %.

Electrical fast transient/burst

IEC 61000-4-4

±2 kV for power supply lines

±1 kV for input/output lines

±2 kV for power supply lines

Not applicable

Mains power quality should be

that of a typical commercial or

hospital environment.

Surge

IEC 61000-4-5

±1 kV differential mode

±2 kV common mode

±1 kV differential mode

±2 kV common mode

Mains power quality should be

that of a typical commercial or

hospital environment.

Voltage dips, short interrup-

tions and voltage variations

on power supply input lines

IEC 61000-4-11

<5 % UT

(>95 % dip in UT)

for 0,5 cycle

40 % UT

(60 % dip in UT)

for 5 cycles

70 % UT

(30 % dip in UT)

for 25 cycles
<5 % UT

(>95 % dip in UT)

for 5 sec

<5 % UT

(>95 % dip in UT)

for 0,5 cycle

40 % UT

(60 % dip in UT)

for 5 cycles

70 % UT

(30 % dip in UT)

for 25 cycles

<5 % UT

(>95 % dip in UT)

for 5 sec

Mains power quality should be

that of a typical commercial

or hospital environment. If

the user of the chair requires

continued operation during

power mains interruptions, it

is recommended that the chair

be powered from an uninter-

ruptible power supply or a

battery.

Power frequency (50/60 Hz)

magnetic field IEC 61000-4-8 3 A / m

3 A / m

Power frequency magnetic

fields should be at levels

characteristic of a typical loca-

tion in a typical commercial or

hospital environment.

NOTE U

T

is the a.c. mains voltage prior to application of the test level.