Ultrasonic milk analyser
Operation manual
23.08.10
35/98
in the milk analyser as Calibration 1 - cow) and press the button Enter.The
following message appears on the display:
Calibration 1
Getting ready
6.4.2.2. Entering values for the separate sample parameters
Wait for the sound signal and the following menu on the display:
Cal1 Samp High
Fat=f.ff
SNF=y.yy
Den=d.dd
------------------------------
Lac=l.ll
Sol=s.ss
Pro=p.pp
Exit
In this display is possible to be entered the results from the milk with high
FAT chemical analyses, which were received by the methods from Appendix
Methods.
For example:
- by using the buttons “up” ▲ and ”down” ▼, choose the needed
parameter to be entered;
- by pressing “Enter” the cursor will mark the number to be changed. For
example: FAT=01.29;
- using the buttons “up” ▲and ”down” ▼, enter the needed value;
- by next pressing of Enter pass towards the next number;
- after completion of entering the needed FAT value, press Enter;
- the cursor returns to its starting position.
Cal 1 Samp.High
Fat=f.ff
SNF=y.yy
Den=d.dd