LABEC M-MIA-CC User Manual
Page 36

Ultrasonic milk analyser
Operation manual
23.08.10
34/98
After releasing the button on the display is shown:
MA Setup
Followed by the possible to be entered by the operator menus:
Special modes
Corrections
Settings
Tests
------------------------------
pH & Co Meter
Accessories
Exit
6.4.1.2. By using buttons “up” ▲ and “down” ▼ position on Settings and
press button Enter.
6.4.1.3. Analogically, position on Recalibrate and press the button Enter.
6.4.2. Making recalibration
6.4.2.1. Determining the calibration mode
After starting Recalibrate, the display shows the following:
Recalibrate
Calibration 1
Calibration 2
Calibration 3
-----------------------------
Edit samp’s 1
Edit samp’s 2
Edit samp’s 3
Edit FrPoints
-----------------------------
Exit
Using the buttons “up” ▲ and “down” ▼ position on the corresponding
calibration (for example if you need to recalibrate cow milk, which is assigned
- AA13 (7 pages)
- GW1060 (44 pages)
- SS1 (4 pages)
- AA-8BS (30 pages)
- AA-22BS (32 pages)
- AA-22N (32 pages)
- AA-8S (33 pages)
- AA-22S (26 pages)
- Shinva-45B (21 pages)
- AA9 (6 pages)
- USC-5L (5 pages)
- WS1.0 (3 pages)
- BEM4 (4 pages)
- LHF-40 (5 pages)
- NBCT2 (4 pages)
- NBCT7S (4 pages)
- ICB-12 (3 pages)
- LBC-4 (4 pages)
- SB-SDC-6 (5 pages)
- J-NBT (16 pages)
- BOB1 (5 pages)
- M-SOD-3636 (3 pages)
- J-SWB35 (21 pages)
- FXP16M (13 pages)
- M-SI-100 (3 pages)
- M-TMO1600 (3 pages)
- M-HOG-020 (13 pages)
- M-HOG010 (12 pages)
- J-USRC (16 pages)
- Z-L-450 (17 pages)
- Z-L-450A (6 pages)
- Z-L-530 (20 pages)
- M-SCEN-206 (13 pages)
- M-SCEN-207 (12 pages)
- Z-ZHWY-2102 (9 pages)
- GLF-90 (2 pages)
- PLCF300 (2 pages)
- H-DW-HW50 (12 pages)
- DW-FL90 (8 pages)
- B-IM0070FSC (5 pages)
- GLR-70 (3 pages)
- PLRGD-1400PT (3 pages)
- PVR-130 (3 pages)
- SPR-100 (3 pages)
- CEMLS (2 pages)