LABEC M-MIA-CC User Manual
Page 27

Ultrasonic milk analyser
Operation manual
23.08.10
25/98
4.2. Preparation of cleaning solution
4.2.1. Preparation of alkaline cleaning solution
Preparation of 1 % alkaline solution of Lactodaily for circulation cleaning in
the milk analyser:
1. Take the package 100 g concentrated powder chemical Lactodaily
2. Carefully cut the upper end, paying attention not to spill it.
3. In appropriate vessel (for example bucket) pour 1 l water.
4. Add the powder and then again water up to 10 l.
Then follow the instruction for milk analyser cleaning.
4.2.2. Preparation of acidic cleaning solution
Preparation of 1 % acidic solution of Lactoweekly for circulation cleaning in
the milk analyser:
1. Take the package 100 g concentrated powder chemical Lactoweekly
2. Carefully cut the upper end, paying attention not to spill it.
3. In appropriate vessel (for example bucket) pour 1 l water.
4. Add the powder and then again water up to 10 l.
Then follow the instruction for milk analyser cleaning.
For final cleaning with the syringe it is recommended to use 1% acidic
solution of Lactoweekly, preliminary heated up to 70 ˚C.
Fig. 8 Labels for acidic cleaning chemical Lactoweekly and alkaline
cleaning chemical Lactodaily