How to carry out linearization:
Call up menu item
Acknowledge with
.
Ensure that there are no objects on the weighing plate.
Wait for stability display, then press
.
When “LoAd 1“ is
displayed, put the first adjustment weight (1/3 max) carefully
in the centre of the weighing platform.
Wait for stability display, then press
. .
When “LoAd 2“
is displayed, put the second adjustment weight (2/3 max)
carefully in the centre of the weighing platform.
Wait for stability display, then press
.
When “LoAd 3“ is
displayed, put the third adjustment weight (1/3 max) carefully
in the centre of the weighing platform.
Wait for stability display, then press
.
After the adjustment the balance will carry out a self-test.
Err
4 will appear briefly, followed by weighing value. The system
is now in weighing mode.
Remove weight, „0“ is displayed.
An adjusting error or incorrect adjusting weight will be
indicated by the error message; repeat adjustment
procedure.
After the linearization the weighing system should be adjusted as
follows: