FlintWeigh Technical Manual, Rev. 3.03 November 2011

Page 27 of 32

8.6. Obligations with Legal for Trade Scales

The operator is responsible for the first and all regularly recurrent official calibrations.

Important: The official seal may only be broken by authorized qualified staff.

If one or more of the following changes has happened, the operater has to send a written note to the assigned
public authority and a new official calibration is mandatory:

Broken seals

Replaced load cells

Replaced load carrier

Replaced wiring

Replaced LDUs

Replaced weighing system controller

Replaced descriptive plate

Calibration of the scale

Changes to the approved software settings


Hint: The user is responsible for the correct performance of any official calibration.

8.7. Legal for Trade Data within the Scale Interface

In legal for trade and industrial applications all calibration data will be stored in a flash memory in the scale
interface. These data can only be accessed, if the authentification of the FlintWeigh II Software to the scale
interface is valid. If these data will be saved during a calibration, the electronic calibration counter (TAC) will
automatically increased by one. The stored data is the following:

Structure of the general scale data
Serial no.

Up to 17x per controller; up to 16x type RC3D digital load cell respective LDU

Gravitational factor

Default value: 9.8075656

Calibration counter

Programme version

x._

Programme revision

_.x

Corner correction

Max. 16 data records for up to 16x type RC3D digital load cell

Checksum EXE

CRC16 for complete programme code

Checksum

CRC16 for general scale data



Structure of the individual scales (3x for Scale 1, Scale 2 and Dual scale)
Scale interval

3x for up to three ranges

Decimals

Unit

Min. capacity

Default value: 20e

No-motion

Calibration weight

Max. capacity

Max. capacity + 9e = overload

Changeover threshold

2x per scale (up to 3 ranges per scale)

Curve linearisation

Up to 5 points per scale

Zero point (Dead load pt.)

Gain value

Scale type

Single-interval, Multi-range, Multi-interval

Checksum

CRC16 for scale data



Logbook
Logbook entries

Up to 200; 8 bytes incl. CRC for each