Preparation for Use
Connecting and Using the AC Adapter (All models)
Connect AC ADAPTER PLUG to AC ADAPTER JACK
Plug AC ADAPTER in the power socket.
AC INDICATOR is lit when AC ADAPTER is connected correctly.
Installing the RECHARGEABLE NiMH Battery (All models)
NOTE: The battery enclosed in the product package is not charged, and needs to be
charged for about 4 hours before use.
NOTE: Turn off the power and unplug AC ADAPTER
before inserting the battery. Make sure "
O " side of
MAIN POWER SWITCH is pressed down.
Remove BATTERY COMPARTMENT COVER by
loosening and removing the screw.
(Fig. 1)
Connect BATTERY PLUG to BATTERY
CONNECTOR inside the battery compartment
and insert the battery.
(Fig. 2)
Put back BATTERY COMPARTMENT COVER and
close it by inserting and tightening the screw.
Connect AC ADAPTER PLUG and plug AC
ADAPTER in the power socket to charge the
battery. The battery is charged regardless of whether
the e-sphyg
TM
2 is powered on or off.
Orange light of AC INDICATOR indicates that
the battery is being charged.
(Fig. 3)
The battery is fully charged after approximately 4 hours.
Unplug AC ADAPTER after the light turns
green as it indicates the battery is charged.
The color of the AC INDICATOR
may switch to green and return to
orange if the 9002 is operated
while charging battery, however it
does not indicate a problem.
(Fig. 1)
(Fig. 2)
(Fig. 3)
ADAPTER
PLUG
BATTERY PLUG
Error Displays and Troubleshooting
If you cannot get correct measurement with the methods above, do not tamper
with the internal mechanism. Contact your dealer. If the unit malfunctions,
return it to the dealer or an authorized service representative for servicing
according to the warranty. The manufacturer provides service data to qualified
dealers.
ERROR SYMBOL/SYMPTOM
CAUSE
REMEDY
Measurement result
Blood pressure was taken in
Measurement result is saved
cannot be recalled or
“MANUAL” mode.
only when it is taken in
is not stored.
“AUTO” mode.
Measurement cannot be
Cuff inflation does
Battery is exhausted.
made when BATTERY MARK
not start.
is displayed. Recharge the
battery.
AC INDICATOR is lit
AC INDICATOR also indicates
even when 9002 is
The battery is inserted.
whether the battery is being
turned off.
charged or it is not being
charged.
Difference in voltages
The column LCD is
between AC ADAPTER and
dim or brightness is
9002 is operated with
the battery lead to differences
not steady.
the battery.
in LCD brightness. This does
not indicate a functional
problem.