Preparation for Use

Connecting and Using the AC Adapter (All models)

Connect AC ADAPTER PLUG to AC ADAPTER JACK

Plug AC ADAPTER in the power socket.

AC INDICATOR is lit when AC ADAPTER is connected correctly.

Installing the RECHARGEABLE NiMH Battery (All models)

NOTE: The battery enclosed in the product package is not charged, and needs to be

charged for about 4 hours before use.

NOTE: Turn off the power and unplug AC ADAPTER

before inserting the battery. Make sure "

O " side of

MAIN POWER SWITCH is pressed down.

Remove BATTERY COMPARTMENT COVER by

loosening and removing the screw.

(Fig. 1)

Connect BATTERY PLUG to BATTERY

CONNECTOR inside the battery compartment

and insert the battery.

(Fig. 2)

Put back BATTERY COMPARTMENT COVER and

close it by inserting and tightening the screw.

Connect AC ADAPTER PLUG and plug AC

ADAPTER in the power socket to charge the

battery. The battery is charged regardless of whether

the e-sphyg

TM

2 is powered on or off.

Orange light of AC INDICATOR indicates that

the battery is being charged.

(Fig. 3)

The battery is fully charged after approximately 4 hours.

Unplug AC ADAPTER after the light turns

green as it indicates the battery is charged.

The color of the AC INDICATOR
may switch to green and return to
orange if the 9002 is operated
while charging battery, however it
does not indicate a problem.

(Fig. 1)

(Fig. 2)

(Fig. 3)

ADAPTER

PLUG

BATTERY PLUG

Error Displays and Troubleshooting

If you cannot get correct measurement with the methods above, do not tamper

with the internal mechanism. Contact your dealer. If the unit malfunctions,

return it to the dealer or an authorized service representative for servicing

according to the warranty. The manufacturer provides service data to qualified

dealers.

ERROR SYMBOL/SYMPTOM

CAUSE

REMEDY

Measurement result

Blood pressure was taken in

Measurement result is saved

cannot be recalled or

“MANUAL” mode.

only when it is taken in

is not stored.

“AUTO” mode.

Measurement cannot be

Cuff inflation does

Battery is exhausted.

made when BATTERY MARK

not start.

is displayed. Recharge the

battery.

AC INDICATOR is lit

AC INDICATOR also indicates

even when 9002 is

The battery is inserted.

whether the battery is being

turned off.

charged or it is not being

charged.

Difference in voltages

The column LCD is

between AC ADAPTER and

dim or brightness is

9002 is operated with

the battery lead to differences

not steady.

the battery.

in LCD brightness. This does

not indicate a functional

problem.