Name and function of each part, Care and maintenance, Cuff sizes – American Diagnostic Corporation (ADC) e-sphyg2 9002 User Manual
Page 8

8
21
Name and function of each part
SMALL ADULT
Arm circumferences of 7.4”-10.6”
(19.0-27.0cm)
ADULT
Arm circumferences of 9.0”-15.7”
(23.0-40.0cm)
LARGE ADULT
Arm circumferences of 13.3”-19.6”
(34.0-50.0cm)
AC ADAPTER
RECHARGEABLE
NICKEL METAL HYBRID
(NiMH) BATTERY
INSTRUCTION
MANUAL
ADAPTER PLUG
BATTERY PLUG
Cuff Sizes
AIR
CONNECTOR
(to main unit)
ADCUFF
TM
BLADDER TUBE
(Female Luer Connection)
COILED TUBING
LUER CONNECTOR
(Male Luer to Female
Bladder Tube)
Because the device includes precision parts, care should be taken to avoid
extreme temperature variations, humidity, shock, dust, and direct sunlight. Do
not drop or strike the device. Make sure not to expose the unit to moisture. This
unit is not water resistant.
Use only a soft, dry cloth to clean the device. Do not use gasoline, paint thin-
ner, chemicals such as strong bases, strong acids, oxidizing agents, and reduc-
ing agents, or other strong solvents on the unit. Since the cuffs may absorb per-
spiration and other fluids, inspect them for stains and discoloration after each
use. Sponge with a damp cloth. If necessary the cuff can be washed in cool
water with mild soap or detergent-disinfectant. Remember to remove the blad-
der prior to washing. After washing, the cuff should be rinsed and allowed to air
dry. NEVER iron.
When storing the device, do not place heavy objects on it and do not coil BLADDER
TUBE too tightly. When the unit has been stored at a temperature below the
freezing point, keep it for at least 1 hour in a warm place before using. Remove
the battery if the device is to be stored for an extended period of time. Keep the
batteries out of the reach of children.
Caution: Do not inflate the cuff when it is not wrapped around an arm.
Caution: Do not disassemble or modify the device.
We suggest that the device be checked every 2 years. This operation may only
be performed by the manufacturer or by firms authorized by the manufacturer.
Checking pressure accuracy
Connect a cuff wrapped around a firm cylinder to the 9002
e-sphyg
TM
2 and a standard manometer. Turn on the 9002
e-sphyg
TM
2 while pressing and holding the START/STOP button
down. Release the button after the initial display (all segment display)
disappears. Two "0"s are displayed and the 9002 e-sphyg
TM
2 is
in CHECKING MODE. Inflate the cuff with an inflation bulb and
compare the readings. Pressure accuracy is to be ±3mmHg.
Contact dealer if accuracy of your 9002 e-sphyg
TM
2 is invalid.
(A test kit may be purchased from an authorized dealer)
9002
e-sphyg
TM
2
9002
PORT
BLADDER
TUBE
Care and Maintenance
Other parts included depending upon model selected.