Name and function of each part

SMALL ADULT

Arm circumferences of 7.4”-10.6”

(19.0-27.0cm)

ADULT

Arm circumferences of 9.0”-15.7”

(23.0-40.0cm)

LARGE ADULT

Arm circumferences of 13.3”-19.6”

(34.0-50.0cm)

AC ADAPTER

RECHARGEABLE

NICKEL METAL HYBRID

(NiMH) BATTERY

INSTRUCTION

MANUAL

ADAPTER PLUG

BATTERY PLUG

Cuff Sizes

AIR

CONNECTOR

(to main unit)

ADCUFF

TM

BLADDER TUBE

(Female Luer Connection)

COILED TUBING

LUER CONNECTOR

(Male Luer to Female

Bladder Tube)

Because the device includes precision parts, care should be taken to avoid

extreme temperature variations, humidity, shock, dust, and direct sunlight. Do

not drop or strike the device. Make sure not to expose the unit to moisture. This

unit is not water resistant.

Use only a soft, dry cloth to clean the device. Do not use gasoline, paint thin-

ner, chemicals such as strong bases, strong acids, oxidizing agents, and reduc-

ing agents, or other strong solvents on the unit. Since the cuffs may absorb per-

spiration and other fluids, inspect them for stains and discoloration after each

use. Sponge with a damp cloth. If necessary the cuff can be washed in cool

water with mild soap or detergent-disinfectant. Remember to remove the blad-

der prior to washing. After washing, the cuff should be rinsed and allowed to air

dry. NEVER iron.

When storing the device, do not place heavy objects on it and do not coil BLADDER

TUBE too tightly. When the unit has been stored at a temperature below the

freezing point, keep it for at least 1 hour in a warm place before using. Remove

the battery if the device is to be stored for an extended period of time. Keep the

batteries out of the reach of children.

Caution: Do not inflate the cuff when it is not wrapped around an arm.

Caution: Do not disassemble or modify the device.

We suggest that the device be checked every 2 years. This operation may only

be performed by the manufacturer or by firms authorized by the manufacturer.

Checking pressure accuracy

Connect a cuff wrapped around a firm cylinder to the 9002

e-sphyg

TM

2 and a standard manometer. Turn on the 9002

e-sphyg

TM

2 while pressing and holding the START/STOP button

down. Release the button after the initial display (all segment display)

disappears. Two "0"s are displayed and the 9002 e-sphyg

TM

2 is

in CHECKING MODE. Inflate the cuff with an inflation bulb and

compare the readings. Pressure accuracy is to be ±3mmHg.

Contact dealer if accuracy of your 9002 e-sphyg

TM

2 is invalid.

(A test kit may be purchased from an authorized dealer)

9002

e-sphyg

TM

2

9002

PORT

BLADDER

TUBE

Care and Maintenance