Error displays and troubleshooting – American Diagnostic Corporation (ADC) e-sphyg2 9002 User Manual

Page 10

background image

10

19

BATTERY MARK flashes in the segment LCD when the battery charge is low.

Only a few more measurements remain before the battery runs down.

The battery needs to be recharged when BATTERY MARK is displayed.

Measurement cannot be made.

NOTE: The rechargeable batteries may not be completely full after their initial charge or when

they have not been used for an extended period of time. This should be solved once the unit

has been used a few times and the batteries have been charged and discharged.

Battery life
Life of the rechargeable NiMH battery is approximately 2 years. However it may vary
depending on operating conditions. Replace the battery when BATTERY MARK is displayed
often or when frequent charging of the battery becomes necessary.

Tips on prolonging the battery life

The device is powered by the AC ADAPTER when it is connected;
the installed battery is not used if the AC ADAPTER is connected.

• Remove the battery when operating the device with the AC ADAPTER for an

extended period of time or when the battery does not need to be charged.

• Avoid charging the battery if it is not completely exhausted.

Do not forcefully remove the battery or pull out the plugs holding
cords and cables.

Use indicated AC Adapter and the battery only.
Contact your dealer if they are damaged.

POWER MARK

The device goes to energy saver mode when it is left turned on and not operated

for approximately 3 minutes. The mark will be displayed for an hour and then disap-

pear. Make sure to turn off the device by pressing "

O " side of MAIN POWER SWITCH

down after each use especially when the device is powered by the battery.

The device activates from energy saver mode when either START/STOP button or

SET/MEMORY button is pressed. It is recommended to activate the device by pressing

SET/MEMORY button since the device could start inflation if START/STOP button is

pressed. Press START/STOP button again and stop inflation if the cuff is not wrapped

around arm.

Error Displays and Troubleshooting

ERROR SYMBOL/SYMPTOM

CAUSE

REMEDY

LOW BATTERY:

Change the battery. Only a

Battery is weak.

few measurements can be

made before the battery

runs down.

BATTERY ERROR:

Charge the battery.

Battery is exhausted.

Measurement cannot be

conducted.

Turn off the device or press

ENERGY SAVER MODE:

START/STOP button or

The device was left turned

SET/MEMORY button to

on and has not been

activate the device from

operated for over

energy saving mode.

approximately 3 minutes.

Make sure to turn off the

device by pressing “

o

” side

of MAIN POWER SWITCH

after use.

Measurement is

When the preset pressure

This does not indicate a

interrupted once and

value was not high enough to

problem.

the cuff is inflated again.

make measurement or when

noise was detected, the cuff

is pressurized again to a

higher level.

Patient moved during

Make sure that the patient

measurement.

stays still and quiet during

measurement.

Patient is not being measured

Blood pressure is

with correct posture.

Make sure that

different each time.

Blood pressure readings

the patient is

The reading is extremely

constantly vary with time

being measured with

low (or high).

of measurement and physical

correct posture.

and mental condition.

Patient moved during

Make sure that the patient

Pulse rate is too low

measurement. There was not

stays still during

(or too high).

enough rest taken before

measurement. Allow your

measurement.

patient to rest for at least

5 minutes.

The indication appearing

during LCD segment test,
which is performed when

the unit is turned on, does

not indicate low battery

or battery error.