Introduction and Intended Use

The 9002 digital ADC

®

e-sphyg

TM

2 sphygmomanometer is a digital

sphygmomanometer intended for measurement of systolic and diastolic

blood pressure and pulse rate in adult patients, i.e., age 12 and above.

This product is not designed for neonatal use. Inaccurate readings may

result if it is used on a child's arm.

Accuracy was evaluated during clinical studies in which results with the

9002 digital ADC

®

e-sphyg

TM

2 sphygmomanometer were compared

with simultaneous auscultatory measurements obtained with a standard

sphygmomanometer. In these studies, the onset of the fourth (or fifth)

Korotkoff sound was taken as the diastolic pressure for the purpose of

determining overall efficacy. A copy of this study is available from ADC

®

on request. To ensure that accuracy is maintained, compare readings

obtained with the 9002 digital ADC

®

e-sphyg

TM

2 sphygmomanometer

with values measured by a trained observer using a manual

sphygmomanometer at least every 6 months.

Blood pressure measurements determined with this device are equivalent

to those obtained by a trained observer using the cuff stethoscope

auscultatory method, within the limits prescribed by the American

National Standard, Manual, electronic, or automated sphygmomanometers.

The 9002 digital ADC

®

e-sphyg

TM

2 sphygmomanometer has the ability

to function according to its specifications in the presence of common

arrhythmia such as atrial or ventricular premature beats or atrial

fibrillation.

Specifications

Operating Principle:

Oscillometric method*

Indicator:

300 digits column LCD and 10 digits segment LCD

Pressure Indicating Range:

0 to 300 mmHg (cuff pressure)

Measuring Range:

50 to 250 mmHg (systolic)

40 to 180 mmHg (diastolic)

40 to 160 bpm (pulse rate)

Accuracy:

±3 mmHg (cuff pressure)
±5% of reading (pulse rate)

Inflation :

Automatic inflation

Deflation:

Electric control valve

Exhaust:

Automatic exhaust valve

Power Supply:

7.0 VDC AC adapter (MODEL 9605+02225-1UOC) or

4.8 VDC rechargeable NiMH battery

Power Consumption :

14W (max.)

Memory:

single measurement result*

Operating Environment:

+50°F to+104°F, 15 to 90% relative humidity

Storage Environment:

-4°F to+122°F, 15 to 95% relative humidity

Coverage Arm Circumference

Adult Size Cuff:

9.0 to 15.7 inches (23.0 to 40.0 cm)

Large Adult Size Cuff:

13.3 to 19.6 inches (34.0 to 50.0 cm)

Small Adult Size Cuff:

7.4 to 10.6 inches (19.0 to 27.0 cm)

Main Unit

Weight:

Approx. 2.2 lb. [1000 g], without AC adapter or battery

Size:

5.4 x 10.6 x 3.8 in [136 x 266 x 96 mm] (W x D x H)

Specifications are subject to change without notice due to improvements in performance.

* marks only apply to "AUTO" measurement.