Introduction and intended use, Specifications – American Diagnostic Corporation (ADC) e-sphyg2 9002 User Manual
Page 4

4
25
Introduction and Intended Use
The 9002 digital ADC
®
e-sphyg
TM
2 sphygmomanometer is a digital
sphygmomanometer intended for measurement of systolic and diastolic
blood pressure and pulse rate in adult patients, i.e., age 12 and above.
This product is not designed for neonatal use. Inaccurate readings may
result if it is used on a child's arm.
Accuracy was evaluated during clinical studies in which results with the
9002 digital ADC
®
e-sphyg
TM
2 sphygmomanometer were compared
with simultaneous auscultatory measurements obtained with a standard
sphygmomanometer. In these studies, the onset of the fourth (or fifth)
Korotkoff sound was taken as the diastolic pressure for the purpose of
determining overall efficacy. A copy of this study is available from ADC
®
on request. To ensure that accuracy is maintained, compare readings
obtained with the 9002 digital ADC
®
e-sphyg
TM
2 sphygmomanometer
with values measured by a trained observer using a manual
sphygmomanometer at least every 6 months.
Blood pressure measurements determined with this device are equivalent
to those obtained by a trained observer using the cuff stethoscope
auscultatory method, within the limits prescribed by the American
National Standard, Manual, electronic, or automated sphygmomanometers.
The 9002 digital ADC
®
e-sphyg
TM
2 sphygmomanometer has the ability
to function according to its specifications in the presence of common
arrhythmia such as atrial or ventricular premature beats or atrial
fibrillation.
Specifications
Operating Principle:
Oscillometric method*
Indicator:
300 digits column LCD and 10 digits segment LCD
Pressure Indicating Range:
0 to 300 mmHg (cuff pressure)
Measuring Range:
50 to 250 mmHg (systolic)
40 to 180 mmHg (diastolic)
40 to 160 bpm (pulse rate)
Accuracy:
±3 mmHg (cuff pressure)
±5% of reading (pulse rate)
Inflation :
Automatic inflation
Deflation:
Electric control valve
Exhaust:
Automatic exhaust valve
Power Supply:
7.0 VDC AC adapter (MODEL 9605+02225-1UOC) or
4.8 VDC rechargeable NiMH battery
Power Consumption :
14W (max.)
Memory:
single measurement result*
Operating Environment:
+50°F to+104°F, 15 to 90% relative humidity
Storage Environment:
-4°F to+122°F, 15 to 95% relative humidity
Coverage Arm Circumference
Adult Size Cuff:
9.0 to 15.7 inches (23.0 to 40.0 cm)
Large Adult Size Cuff:
13.3 to 19.6 inches (34.0 to 50.0 cm)
Small Adult Size Cuff:
7.4 to 10.6 inches (19.0 to 27.0 cm)
Main Unit
Weight:
Approx. 2.2 lb. [1000 g], without AC adapter or battery
Size:
5.4 x 10.6 x 3.8 in [136 x 266 x 96 mm] (W x D x H)
Specifications are subject to change without notice due to improvements in performance.
* marks only apply to "AUTO" measurement.