Resetting to default settings – ALTANA LC 3 User Manual
Page 11
![background image](/manuals/659698/11/background.png)
BYK LC 3
Resetting to default settings
11
Resetting to default settings
You can reset the measurement parameters and the configuration to the de-
livery status separately from one another (initialization).
Measurement
parameters
The following measurement parameters can be reset to the delivery status:
Configuration
parameters
The following configuration parameters (InI) can be reset to the delivery sta-
tus:
Reset procedure:
Conductivity (
InI)
Measuring mode
AutoRange, automatic switchover
of the measurement range
On (YES)
Cell constant
0.475 cm
-1
(calibrated)
0.475 cm
-1
(manual range I)
0.100 cm
-1
(manual range II)
Temperature compensation
nLF
Temperature coefficient of the lin-
ear temperature compensation
2.000 %/K
TDS factor
1.00
Baud rate
4800
Interval 1
(automatically saved)
OFF
Interval 2
(for data transmission)
OFF
! To switch it off, press
! To reset the measurement parameters with <
▲> <▼>, select YES and
confirm with
! Continue to the configuration parameters (InI) without resetting with the
! After the configuration parameters, InI changes the instrument to the last
active measurement mode.
o
n
I
n
I