beautypg.com

3 flame detector, 4 combustion calibration, 5 safety device testing – AERCO BMK 2.0 LN Natural Gas 2008 User Manual

Page 50: Maintenance

background image

MAINTENANCE

Table 7-1 - Maintenance Schedule

PARAGRAPH

ITEM

6 Mos.

12 Mos.

24 Mos.

Labor

Time

7.2

Spark Igniter

(GP-122435-S)

*Inspect Inspect Replace 15

mins.

7.3

Flame Detector

(66006)

*Inspect Inspect Replace 15

mins.

7.4

Combustion

Calibration

*Check Check

1

hr.

7.5

Testing of

Safety Devices

See CSD-1

Chart in

Appendix I

20

mins.

7.6 Burner Inspect

2

hrs.

7.7

Condensate

Drain Trap

*Inspect

Inspect &

Clean

15

mins.

* Only performed after initial 6 month period after initial startup.

7.3 FLAME DETECTOR

The flame detector (part no. 66006) is located
on the burner plate (see Figure 7-1). The flame
detector may be HOT. Allow the unit to cool
sufficiently before removing the flame detector.

To inspect or replace the flame detector:

1. Set the ON/OFF switch on the control panel,

to the OFF position. Disconnect AC power
from the unit.

2. Remove the top panels from the unit.

3. Disconnect the flame detector lead wire.

Unscrew the flame detector and remove it.
(See Figure 7-2).

4. Inspect the detector thoroughly. If eroded,

the detector should be replaced. Otherwise
clean the detector with a fine emery cloth.

5. Reinstall the flame detector and flame

detector gasket, if removed.

6. Reconnect the flame detector lead wire.

7. Reinstall the side and top panels on the unit.

7.4 COMBUSTION CALIBRATION

Combustion settings must be checked at the
intervals shown in Table 1 as part of the
maintenance requirements. Refer to Chapter 4
for combustion calibration instructions.

AIR/FUEL

VALVE

BLOWER

BURNER

PLATE

SPARK

IGNITOR

BURNER

FLAME

DETECTOR

BLOCKED

INLET

SWITCH

BLOWER

PROOF

SWITCH

Figure 7-2

Burner Assembly Spark Igniter and

Flame Detector Location

7.5 SAFETY DEVICE TESTING

Systematic and thorough tests of the operating
and safety devices should be performed to
ensure that they are operating as designed.
Certain code requirements, such as ASME
CSD-1, require that these tests be performed on
a scheduled basis. Test schedules must
conform to local jurisdictions. The results of the
tests should be recorded in a log book. See
Chapter 6-Safety Device Testing Procedures.

7-2