13. APPENDIX

The table below lists the relation between the ISO 8655 requirements and the PT series pipette

accuracy testers.

ISO 8655 requirements

Maximum permissible error

Pipette
nominal
volume

*4

Accuracy Repeatability

Balance

minimum

weighing

value

(

μL) ±% ±μL % μL mg

Pipette accuracy tester
corresponding to the requirements

20 1.0

0.2

0.5

0.1

50 1.0

0.5

0.4

0.2

100 0.8 0.8

0.3 0.3

0.01 AD-4212B-PT

*5

200 0.8 1.6

0.3 0.6

500 0.8 4.0

0.3 1.5

1000 0.8 8.0 0.3 3.0
2000 0.8 16 0.3 6.0
5000 0.8 40 0.3 15.0

10000 0.6 60 0.3 30.0

0.1

AD-4212A-PT

Daily inspection, simplified verification

1

*6

FX-300i-PT

*4

The maximum volume selectable for adjustable volume pipettes

*5

The AD-4212B-PT can be used for the pipette volume range from 20

μL to 10000 μL.

*6

The minimum weighing value, 1 mg, approximately corresponds to 1

μL. If a pipette volume is

1000

μL, a test can be performed with a resolution of 0.1%. If 200 μL, 0.5%.


NOTE Pipette accuracy testing is performed using a high-resolution valance. Make sure that

the measurement environment is free from vibration, drafts and air from air
conditioners. For details, refer to “3. PRECAUTIONS BEFORE USE”.