Calibrating the balance – A&D Weighing FX-300i-PT User Manual
Page 16
14
7. CALIBRATING THE BALANCE
For details on calibration, read the balance instruction manual in addition to this manual.
Before calibration, refer to “3. PRECAUTIONS BEFORE USE”. Especially when calibrating the
AD-4212B-PT / AD-4212A-PT with a minimum weighing value of 0.01 mg / 0.1 mg, make sure that the
installation environment is appropriate and the balance is warmed up for the appropriate duration.
7-1 AD-4212B-PT / AD-4212A-PT
1.
Remove the sample cup, but leave two
sample cup holders (5 mL and 30 mL) inside
the evaporation trap, as shown to the right.
2. Press and hold the CAL key until Calout
is displayed, then release the CAL key.
3. When Cal 0 is displayed, press the
PRINT key to enter the zero point.
4. After the zero point is entered, the calibration
weight value 50 is displayed.
Remove the upper part of the evaporation
trap while holding the lower part of the
evaporation trap.
Place the accessory calibration weight on the
sample cup holder (5 mL).
Attach the upper part of the evaporation trap
to the lower part, mating the tab with the
notch.
5.
Press the PRINT key to calibrate the
weighing capacity. When end is
displayed, remove the calibration weight. The
balance will return to the weighing mode.
6. Place the calibration weight on the
sample
cup holder (5 mL) to confirm that calibration
has been completed successfully.
Calibration
weight (50 g)
Tab
Notch
Upper part of
the evaporation trap
Sample cup
holder (30 mL)
Lower part of
the evaporation trap
Sample cup
holder (5 mL)