10. AD-4212B-PT / AD-4212A-PT


The AD-4212B-PT / AD-4212A-PT is designed exclusively as a pipette accuracy tester.

Although the performance as a balance is the same as the AD-4212B / AD-4212A production

weighing unit, some of the specifications are different.

10-1 Factory Settings

When shipped, the factory settings of the AD-4212B-PT / AD-4212A-PT as a pipette accuracy tester

are as follows.

Unit

z

Weighing unit: mg

z

Minimum weighing value for the AD-4212B-PT: 0.1 mg

Press the SAMPLE key to switch to 0.01 mg as necessary, particularly for testing a liquid with a

volume of less than 100

μL.

Function table

Functions related to pipette accuracy testing are as follows.

Class

Item and Parameter

Description

0

1

Cond

Condition

2

Fast weighing speed, sensitive value

FAST

MID.
Slow weighing speed, stable value

SLOW

0

OFF

1

Normal

2

Strong

trc

Zero tracking

3

Very strong

0

5 times/second

5pd

Display refresh rate

1

10 times/second

0

OFF

ba5fnc

Environment
Display

ploc

Display lock function

1

ON

0

No comparison

1

Comparison, excluding “near zero” when stable or overloaded

2

Comparison, including “near zero” when stable or overloaded

3

Continuous comparison, excluding “near zero”

Cp fnc

Comparator

Cp

Comparator mode

4

Continuous comparison, including “near zero”

Factory setting