7-2 FX-300i-PT

1. Remove the 30 mL sample cup and leave

only the

sample cup holder (30 mL) inside

the evaporation trap as shown to the right.

2. Press and hold the CAL key until Calout

is displayed, then release the CAL key.

3. When Cal 0 is displayed, press the

PRINT key to enter the zero point.

4. After the zero point is entered, the calibration

weight value 100 is displayed.

Remove the evaporation trap by holding the

lower part of the evaporation trap.

Place the accessory calibration weight on the

sample cup holder (30 mL).

Attach the evaporation trap on the base,

mating the tabs with the holes.

5.

Press the PRINT key to calibrate the

weighing capacity. When end is

displayed, remove the calibration weight. The

balance will return to the weighing mode.

6. Place the calibration weight on the

sample

cup holder (30 mL) to confirm that calibration

has been completed successfully.


NOTE The evaporation trap can be used as a breeze break.

Calibration can be performed, using the weighing pan and the breeze break provided
as accessories.


Upper part of
the evaporation trap


Lower part of
the evaporation trap

Sample cup
holder (30 mL)

Calibration
weight (100 g)

Tab
(4 in all)

Hole
(4 in all)

Base