A&D Weighing SV-100 User Manual
Page 46
44
Output format example (1) With ID number, date and time added
Function setting
(
√=Output Blank=No output)
5-td 1 Date/time
√
5-ed 3
Device
ID number
√
SV-10 output format example
Viscosity
/
Temper-
ature
Display
Output format example
Remarks
L
LLLL
mPa・s
LAB-12,2003/03/19,12:34:56,+025.67,C,+00000.00,mPa s
Zeroes are output for
below measuring
range error.
0.30 mPa・s
LAB-12,2003/03/19,12:34:56,+025.67,C,+00000.30,mPa s
10.0 mPa・s
LAB-12,2003/03/19,12:34:56,+025.67,C,+00010.00,mPa s
100 mPa・s
LAB-12,2003/03/19,12:34:56,+025.67,C,+00100.00,mPa s
1.00 Pa・s
LAB-12,2003/03/19,12:34:56,+025.67,C,+01000.00,mPa s
For 1000 mPa
⋅s or
greater, the displayed
unit is Pa
⋅s, but the output
unit remains mPa
⋅s.
mPa・s
/
°C
H Pa・s
LAB-12,2003/03/19,12:34:56,+025.67,C,+12000.00,mPa s
12000 is output for
above measuring
range error.
L Pa・s
LAB-12,2003/03/19,12:34:56,+051.23,F,+000.0000, Pa s
Zeroes are output for
below measuring
range error.
0.0003 Pa・s
LAB-12,2003/03/19,12:34:56,+051.23,F,+000.0003, Pa s
0.0100 Pa・s
LAB-12,2003/03/19,12:34:56,+051.23,F,+000.0100, Pa s
0.100 Pa・s
LAB-12,2003/03/19,12:34:56,+051.23,F,+000.1000, Pa s
1.00 Pa・s
LAB-12,2003/03/19,12:34:56,+051.23,F,+001.0000, Pa s
Pa・s
/
°F
H Pa・s
LAB-12,2003/03/19,12:34:56,+051.23,F,+012.0000, Pa s
12 is output for above
measuring range error.
L cP
LAB-12,2003/03/19,12:34:56,+025.67,C,+000.0000, cP
Zeroes are output for
below measuring
range error.
0.30 cP
LAB-12,2003/03/19,12:34:56,+025.67,C,+00000.30, cP
10.0 cP
LAB-12,2003/03/19,12:34:56,+025.67,C,+00010.00, cP
100
cP
LAB-12,2003/03/19,12:34:56,+025.67,C,+00100.00, cP
10.0 P
LAB-12,2003/03/19,12:34:56,+025.67,C,+01000.00, cP
For 1000 cP or greater,
the displayed unit is
P, but the output unit
remains cP.
cP
/
°C
H P
LAB-12,2003/03/19,12:34:56,+025.67,C,+12000.00, cP
12000 is output for
above measuring
range error.
L P
LAB-12,2003/03/19,12:34:56,+051.23,F,+000.0000,
P
Zeroes are output for
below measuring
range error.
0.0030 P
LAB-12,2003/03/19,12:34:56,+051.23,F,+000.0030,
P
0.100 P
LAB-12,2003/03/19,12:34:56,+051.23,F,+000.1000,
P
1.00 P
LAB-12,2003/03/19,12:34:56,+051.23,F,+001.0000,
P
10.0 P
LAB-12,2003/03/19,12:34:56,+051.23,F,+010.0000,
P
P
/
°F
H P
LAB-12,2003/03/19,12:34:56,+051.23,F,+120.0000,
P
120 is output for above
measuring range error.
: Space (ASC 20h)
Outputs in the order of ID number, date, time,
temperature, temperature unit, viscosity and viscosity
unit.
The output data are 52 characters long excluding the
terminator.