beautypg.com

A&D Weighing SV-100 User Manual

Page 37

background image

35

Decimal Point (pnt)

Parameter

Settings Description

0

• Dot

"

.

"

1

Comma "

,

"

The decimal point format for the displayed measurement data and the
decimal point code for measurement data output via RS-232C are
specified.

With "Comma" selected, the separator for CSV format and RsVisco
format will be "

;

" (semicolon).

MODE Key Function During Measurement (fnc)

Parameter

Settings Description

0

Switches the viscosity
units.

Each time the MODE key is pressed, the viscosity unit is
switched.

SV-10: mPa

⋅s ⇔ Pa⋅s, cP ⇔ P

sv-100: Pa

⋅s ⇔ P

Note
With the SV-10, for the viscosity over 1000 mPa

s, the

unit is fixed to Pa

s and for the viscosity over 1000 cP,

the unit is fixed to P.

1

Switches between the
temperature display
and the measurement
elapsed time display

Each time the MODE key is pressed, the display is switched
between the temperature display and the measurement
elapsed time display.

Upon a measurement start, the temperature display is
selected.

When the elapsed time reaches 100 hours, the display
returns to 0. (99.59.59

→00.00.00)


Data Output Mode (prt)

The condition to output the measurement data via RS-232C is set.

Parameter

Settings Description

0

• Key mode

During measurement or in the data hold mode, pressing the

PRINT key outputs the current measurement values. *2

1

Auto print mode

The measurement values are output automatically when the

STOP key ends the measurement.

Pressing the PRINT key outputs the current measurement
values. *2

2

Stream mode

The measurement values are output continuously during
measurement.

When D.P. format is selected in "Output format (type)" of the
function setting, only the viscosity value is output, regardless
of the settings of "5-at", "5-td" and "5-ed".

When this mode is selected, the data hold mode using the

HOLD key is not available.

*2 When A&D standard format is selected in "Output format (type)" of the function setting,

pressing the PRINT key in the standby mode will not output the measurement data.