A&D Weighing SV-100 User Manual
Page 41
39
Initialization (Clr)
Restores the following data to the default setting.
Function setting
Calibration data
After initialization, check the viscosity value and perform calibration as necessary. (Refer to
6. VISCOSITY CALIBRATION").
1 In the standby mode, press and hold the
MODE key to enter the function setting
mode. " Cladj " appears.
2 Press the MODE key to select " Clr ".
3 Press the PRINT key to display " Clr no ".
4 Press the START key to select " Clr go ".
5 Press the PRINT key to execute initialization.
After " end ", the next item is displayed.
Initialization has completed.
6 Press the STOP key to return to the standby
mode.
This manual is related to the following products: