4-3 After Use
Remove any residual sample material from the sensor plates, temperature sensor and
protector using alcohol. Using the sensor plates, temperature sensor and protector with
residue of an old sample left on will cause a measurement error.
Clean the sensor plates carefully to avoid bending them.
The sensor plates and the temperature sensor are made of stainless steel (SUS304). The
surface is plated with 24K gold.
Note
Liquids with strong acidity may remove the gold plating and corrode the sensor plates
and the temperature sensor.
How to clean the sensor plates and temperature sensor
Hold the sensor plate or temperature sensor with
tissue paper. Move the tissue paper downward to
remove the sample.
Then, use tissue paper moistened with alcohol,
to remove any residual sample material.
Clean the sample cup as necessary.
Unlock the cable connector before disconnecting the connection cable.
How to unlock the cable connector