A&D Weighing SV-100 User Manual
Page 25
23
6-2-1 One-point Calibration
1 In the standby mode, press and hold
the HOLD key to enter the
calibration mode. " Cal " appears.
2 Select one-point calibration (Cal-1)
and press the PRINT key to confirm.
The standby mode of the one-
calibration mode appears.
Use the MODE key to switch
between the calibration modes,
one-point calibration (Cal-1) or
two-point calibration (Cal-2).
Press the STOP key to exit the
calibration mode. The display returns
to the standby mode.
3 Place the standard viscosity fluid in
the sample cup. Press the START
key to start a measurement.
4 After the measurement, wait for the
display to become stable and press
the PRINT key. The measurement
value blinks and is ready to be
corrected.
5 Correct the value using the following keys:
MODE key
Switches the blinking digits.
START key
Changes the setting of a
blinking digit.
STOP key
Moves the decimal point.
6 Press the PRINT key to confirm the correction value.
7 To exit the calibration mode:
With " end " being displayed, press the PRINT
key again to return to the standby mode.
To correct the value:
With " end " being displayed, press the
STOP key and correct the value.
8 Measure the viscosity of the standard
viscosity fluid used. Confirm that the
viscometer displays the similar value
as the entered correction value.
The below is an example of the SV-10.
For the SV-100, the unit is Pa
⋅s
(Standby mode)
(One-point calibration)
Press and hold
(One-point calibration
standby mode)
(Measurement mode)
When the data
is stabilized
(Correct the value)
(End of one-point calibration)
(Standby mode)
To correct
Set standard viscosity fluid
(Confirm the value)