2 details of the function items – A&D Weighing SV-100 User Manual
Page 32

30
7-2 Details of the Function Items
Function item
Parameter
Description
Cladj
Date/Time
Sets the order of the date (YMD,MDY,DMY) and the
date/time.
Cond 0
Follows the viscosity changes quickly. (Prone to vibration)
Condition
1
•
2
Follows the viscosity changes slowly. (Stable values)
Unit 0
• mPa⋅s
Unit upon power-on
1
• Pa⋅s
2
cP
3
P
°C
4
mPa
⋅s
5
Pa
⋅s
6
cP
7
Viscos
-ity
P
Temper-
ature
°F
Factory setting:
SV-10=0
SV-100=1
pnt 0
• Dot
Decimal point
1
Comma
With "Comma" selected, the
separator for CSV format will
be ";" (semicolon).
fnc 0
• Switches viscosity units.
MODE key function
during measurement
1
Switches between the temperature display and the
measurement elapsed time display.
prt 0
• Key
mode
Press the PRINT key to output data.
Data output mode
1
Auto print mode
Outputs automatically when
the STOP key ends the
measurement.
2
Stream
mode
Continuous output during
measurement. Outputs the viscosity
only when D.P. format is selected.
type 0
A&D standard format
For AD-8121B MODES 1&2
Data output format
1
• D.P. format
For AD-8121B MODE 3
2
CSV format
For a personal computer
3
RsVisco format
For graphing program RsVisco
5-at 0
No
output
Measurement
elapsed time output
1
• Output
Available only for D.P.
format
5-td 0
No
output
Date/time output
1
• Output
Available only for D.P. and
CSV formats
5-ed 0
No
output
Other output 1
Outputs
remarks.
2
• Outputs remarks, Device ID
information and signature.
Available only for D.P. format
3
Outputs ID number.
Available only for CSV format
pU5e 0
No
pause
Pause at data
output
1
• Pause (Approx. 2 seconds)
erfnc 0
• Usually use this parameter.
Reserved
|
7
id
Device ID number
Set the device ID number.
With "5-ed", the device ID information
is added to the measurement data.
Clr
Initialization
Restores the function settings and calibration data to the
factory setting.
• Factory setting