Operation qualification (oq test) – ChemoMetec NC-3000 User Manual
Page 16

991-3003 Rev. 1.0
August 2013
16
Operation Qualification (OQ Test)
1. Start the NucleoView NC-3000™ software by double-clicking the
icon on the desktop.
2. Wait while the NC-3000™ instrument is initializing. During this time, the motors may be heard
positioning.
3. Click the
button (just below the F3 button in the right side of the main window) to launch the
Select Protocol window.
4. In the Select Protocol window choose "Verification of NC-3000" and "Test/calibration of NC-3000".
Select the "OQ Test" protocol in the list box in the lower half of the window (make sure that the
NC-Slide A2 checkbox is checked).
5. Click the button to see the application note for the selected protocol.
Check that you have all accessories needed for this protocol available (Cassette, Slides, beads).
6. Follow the instructions on screen and in the application note to finalize the OQ Test.