beautypg.com

Installation qualification (iq test) – ChemoMetec NC-3000 User Manual

Page 15

background image

991-3003 Rev. 1.0

August 2013

15


Installation Qualification (IQ Test)

1. Start the NucleoView NC-3000™ software by double-clicking the

icon on the desktop.

2. Wait while the NucleoCounter® NC-3000™ initialize. During this time, the motors may be heard

positioning.

3. Click the

button (just below the F3 button in the right side of the main window) to launch the

Select Protocol window.

4. In the Select Protocol window choose "Verification of NC-3000" and "Test/calibration of NC-3000".

Select the "IQ Test" protocol in the list box in the lower half of the window (make sure that the NC-
Slide A2 checkbox is checked).

5. Click the button to see the application note for the selected protocol.
6. Follow the instructions on screen and in the application note to finalize the IQ Test.