6:1:2 ZERO

When installed on the competely slack rope, the RTM should indicate zero on the dis-
play otherwise you have to zeroize the instrument. When you set to zero hold the RTM
with the tightning device hanging vertically or put the RTM in this position on a slack
rope.

The instrument is zeroed by pressing and holding the

ENTER

/ZERO for about 2 seconds.

The display will indicate
This zero-setting remains saved even if the RTM is turned off.
This zero-setting should only be done when the rope is untensioned.

6:1:3 Peak Value - highest displayed value

The instrument stores automaticly the highest displayed value. This peak-value can
originate from any of the different “Rope No”s.
This peak-value remains saved even if the RTM is turned off.

To display peak-value press and hold in the PEAK - button.

When the button is released the display will indicate the present tension value.

Do not forget to reset the Peak Value before doing new measurements.

To reset the peak-value, do the following:

PEAK +

ENTER

/ZERO i.e.

press and hold in the PEAK button
press the

ENTER

/ZERO button too

release the

ENTER

/ZERO and PEAK buttons.

The peak value is also resetted if the instrument is set to zero.